Article Dans Une Revue Journal of Instrumentation Année : 2010

Time Reconstruction and Performance of the CMS Electromagnetic Calorimeter

S. Chatrchyan
  • Fonction : Auteur
Pierre Busson
P. Paganini
J.-L. Agram
E. Conte
D. Contardo
H. El Mamouni
  • Fonction : Auteur

Résumé

The resolution and the linearity of time measurements made with the CMS electromagnetic calorimeter are studied with samples of data from test beam electrons, cosmic rays, and beam-produced muons. The resulting time resolution measured by lead tungstate crystals is better than 100 ps for energy deposits larger than 10 GeV. Crystal-to-crystal synchronization with a precision of 500 ps is performed using muons produced with the first LHC beams in 2008.

Dates et versions

in2p3-00664804 , version 1 (31-01-2012)

Identifiants

Citer

S. Chatrchyan, P. Nedelec, D. Sillou, M. Besancon, R. Chipaux, et al.. Time Reconstruction and Performance of the CMS Electromagnetic Calorimeter. Journal of Instrumentation, 2010, 5(03), pp.T03011. ⟨10.1088/1748-0221/5/03/T03011⟩. ⟨in2p3-00664804⟩