Frequency analysis of the noise in the fowler(n) sampling of a H2RG(2Kx2K) near-IR detector

Résumé

The readout noise of a H2RG HgCdTe NIR detector from Teledyne is measured at a temperature T=110K. It is shown that a Fowler mode with n = 240 allows to reach a noise of 2.63e (single read). A description of the power spectrum in terms of 3 parameters reproduces the variation of the noise as a function the number of Fowler samples, as well as its dependence on the periodicity of the sampling. The variance of the noise decreases with frequency with an effective power of 0.62 in our measurement domain. The behaviour of the detector under different experimental conditions can then be predicted

Dates et versions

in2p3-00498217 , version 1 (07-07-2010)

Identifiants

Citer

G. Smadja, C. Cerna, A. Castera, A. Ealet. Frequency analysis of the noise in the fowler(n) sampling of a H2RG(2Kx2K) near-IR detector. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2010, 622, pp.288-294. ⟨10.1016/j.nima.2010.06.271⟩. ⟨in2p3-00498217⟩