Geant4 simulation of the response of phosphor screens for X-ray imaging

S.A. Pistrui-Maximean
  • Fonction : Auteur
N. Freud
J.M. Létang
A. Koch
  • Fonction : Auteur
B. Munier
  • Fonction : Auteur
A.H. Walenta
  • Fonction : Auteur
D. Babot
  • Fonction : Auteur

Résumé

In order to predict and optimize the response of phosphor screens, it is important to understand the role played by the different physical processes inside the scintillator layer. Monte Carlo simulations were carried out to determine the Modulation Transfer Function (MTF) of phosphor screens for energies used in X-ray medical imaging and nondestructive testing applications. The visualization of the dose distribution inside the phosphor layer gives an insight into how the MTF is progressively degraded by X-ray and electron transport. The simulation model allows to study the influence of physical and technological parameters on the detector performances, as well as to design and optimize new detector configurations. Preliminary MTF measurements have been carried out and agreement with experimental data has been found in the case of a commercial screen (Kodak Lanex Fine), at an X-ray tube potential voltage of 100 kV. Further validation with other screens (transparent or granular) at different energies is under way.
Fichier principal
Vignette du fichier
iworid.pdf (198.75 Ko) Télécharger le fichier
Loading...

Dates et versions

in2p3-00025011 , version 1 (07-11-2005)

Identifiants

Citer

S.A. Pistrui-Maximean, N. Freud, J.M. Létang, A. Koch, B. Munier, et al.. Geant4 simulation of the response of phosphor screens for X-ray imaging. International Workshop on Radiation Imaging Detectors IWORID-7, Jul 2005, Grenoble, France. pp.196-199, ⟨10.1016/j.nima.2006.01.094⟩. ⟨in2p3-00025011⟩
42 Consultations
478 Téléchargements

Altmetric

Partager

More